32 N9030B PXA Signal Analyzer Service Guide
Overview
Service Equipment You Will Need
Required Test Equipment List
The following table identifies the equipment recommended for
troubleshooting, adjusting, and verifying the performance of the instrument.
Only the recommended and alternate equipment is compatible with the
performance verification testing. Some tests can use various models of a
particular equipment type. The “Recommended Keysight Model” is the
preferred equipment. However, the “Alternative Keysight Model” is an
acceptable substitute.
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a
Use
b
Signal Sources
Microwave Signal
Generator #1
Frequency: 10 MHz to 50 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −30 dBc
Amplitude range: -40 dBm to +13 dBm
Amplitude resolution: 0.02
VSWR: < 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
E8257D
(Option 567, 1EA/1EU,
1E1) (Option 007,
Analog Ramp Sweep,
required for Frequency
Response
Adjustments)
For
Adjustments:
83630A/B/L
(Option 001,
008),
83640A/B/L
(Option 001,
008),
83650A/B/L
(Option 001,
008),
PSG
c
For
Performance
Tests:
83630A/B
(Option 001,
008),
83640A/B
(Option 001,
008),
83650A/B
(Option 001,
008)
PSG
c
A, P, T