N9030B PXA Signal Analyzer Service Guide 33
Overview
Service Equipment You Will Need
Microwave Signal
Generator #2
(only required for Third
Order Intermodulation and
Gain Compression)
Frequency: 10 MHz to 26.5 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −30 dBc
Amplitude range: -20 dBm to +13
Amplitude resolution: 0.02
VSWR: < 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
E8257D
(Option 1EA/1EU)
83630A/B
(Option 008),
83640A/B
(Option 008),
83650A/B
(Option 008)
PSG
c
P
Low Noise
Signal Generator
(one PSG may be used for
multiple sources.
See footnote
c
)
Frequency: 50 MHz to 1.0 GHz
Harmonics: < -30 dBc ≤ +13 dBm
Spectral Purity SSB Phase Noise @ 1
GHz:
−112 dBc at 100 Hz offset
−121 dBc at 1 kHz offset
−131 dBc at 10 kHz offset
VSWR: < 1.5:1
E8257D
(Options UNR, UNX, or
UNY)
8663A
PSG
c
(Options UNR,
UNX, or UNY)
A, P, T
RF Signal Generator
(one PSG may be used for
multiple sources.
See footnote
c
)
Frequency: 300 kHz to 3.6 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −40 dBc
Amplitude range: -20 dBm to +13 dBm
Amplitude resolution: 0.02
Harmonic level:
< 2 GHz, < -30 dBc
≥ 2 GHz & < 26.5 GHz, < -50 dBc
VSWR: < 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
For Option UNR or UNX:
Harmonics: < -30 dBc ≤ +13 dBm
Spectral Purity SSB Phase Noise @ 1
GHz:
−112 dBc at 100 Hz offset
−121 dBc at 1 kHz offset
−131 dBc at 10 kHz offset
VSWR: < 1.5:1
E8257D
(Options UNR, UNX, or
UNY)
8665A/B
(for Frequency
Response
below 3.6 GHz
test and Phase
Noise > 30 kHz
test only)
PSG
c
(Options UNR,
UNX, or UNY)
A, P
Ultra Low Noise Reference
Frequency Source
(required for testing PXA
Phase Noise)
Phase Noise (dbc/Hz) for 1 GHz tone and
these offsets:
100 Hz: −104 dBc/Hz
1 kHz: −132 dBc/Hz
10 kHz: −144 dBc/Hz
30 kHz: −146 dBc/Hz
100 kHz: −148 dBc/Hz
1 MHz: −157 dBc/Hz
10 MHz: −164 dBc/Hz
Wenzel Associates
Inc
d
Ultra Low Noise
Reference Frequency
Source
Model 500-13438
Rev D
P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a
Use
b