34 N9030B PXA Signal Analyzer Service Guide
Overview
Service Equipment You Will Need
Function Generator Frequency: 3 Hz to 300 kHz
Amplitude Resolution: 0.1 mv
Harmonic Distortion: -35 dBc
33250A
(required for Channel
Match (Option BBA)
test)
33120A
(Option 001)
(cannot be used
for CXA
frequency
response
adjustment)
A, P
Counters
Universal Counter Frequency: 10 MHz
Gate time: 10 to 100 seconds
Must be capable of measuring signal at
+7 dBm (0.5 Vrms)
53132A 53131A A, P, T
Meters
Digital Multimeter Frequency: 3 Hz to 300 kHz
AC Accuracy: ± 0.31% of reading
3458A A, P, T
Power Meter Dual Channel
Instrumentation Accuracy: ± 0.5%
Power Reference Accuracy: ± 0.6%
Compatible with 8480 series power
sensors
dB relative mode
N1914A E4419A/B
N1912A
A, P
RF Power Sensor
(2 required)
Frequency Range: 100 kHz to 3.6 GHz
Amplitude Range: −30 to +20 dBm
Zero Set: ± 50 mW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
VSWR 100 kHz to 1 MHz: ≤ 1.20:1
1 MHz to 2 GHz: ≤ 1.10:1
2 GHz to 3 GHz: ≤ 1.30:1
Cal Factor Uncertainty (std): < 1.6%
Option H84 Cal Factor: Characterized by
standards lab to: ± 0.6%
e
Input Connector: Type-N (m)
8482A
(Option H84)
N8482A
f
N8482A CFT
f
8482A
f
(Non H84
sensors will
increase
measurement
uncertainty)
A, P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a
Use
b