N9030B PXA Signal Analyzer Service Guide 35
Overview
Service Equipment You Will Need
Microwave Power Sensor
(2 required)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: -30 to +20 dB
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
N8485A
f
N8485A CFT
f
8485A
f
A, P
Millimeter Power Sensor
(for Options > 526)
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: 30 to +20 dBm
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 100 nW
Cal Factor Uncertainty: < 4.5%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
26.5 GHz to 40 GHz: 1.30:1
40 GHz to 50 GHz: 1.50:1
Input Connector: 2.4 mm coaxial (m)
N8487A 8487A A, P
High Sensitivity Millimeter
Power Sensor
(for Options > 526)
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: 70 to 20 dBm
Zero Set: ± 20 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty (std): < 4.5%
VSWR: 50 MHz to 100 MHz: 1.19:1
100 MHz to 4 GHz: 1.15:1
4 GHz to 12.4 GHz: 1.20:11
2.4 GHz to 18 GHz: 1.29:1
18 GHz to 34 GHz: 1.37:1
34 GHz to 40 GHz: 1.61:1
40 GHz to 50 GHz: 1.89:1
Input Connector: 2.4 mm (m)
8487D A, P
Table 1-1 Required Test Equipment
Instrument Critical Specifications Recommended
Keysight Model
Number
a
or Vendor
Model Number
Alternative
Keysight Model
Number
a