N9030B PXA Signal Analyzer Service Guide 215
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Even if the voltages listed in the table are correct it does not guarantee the
proper control current is getting to the YTF. There is not a convenient way to
measure the control current, so if a high band issue is suspect it is
recommended that the A15 is changed out first before changing the YTF since
the A15 board has easy access.
Table 6-12 Preselector Tune Output Voltages (Option 544, 550)
Center Frequency
(GHz)
~ Tune Voltage
(VDC)
Valid Frequency
Range Option
5.0 0.9 544, 550
10.0 1.8 544, 550
15.0 2.7 544, 550
20.0 3.6 544, 550
25.0 4.6 544, 550
30.0 5.5 544, 550
40.0 7.3 544, 550
42.5 7.75 544, 550
43.5 8.0 544, 550
45 8.25 550
50 9.2 550