538 N9030B PXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Input Connector Assembly
2. Reconnect the semi-rigid cable W2 to the Input Connector Assembly.
Torque to 10 inch-pounds.
3. Replace the Front Frame Assembly. Refer to the Front Frame Assembly
replacement procedure.
4. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.