108 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–221 Settings conflict; Trace
Math is not available while
Normalize is on
The Normalize function works by doing trace manipulation,
so trace math is not available while normalization is running.
–221 Settings conflict; Tracking
Source unavailable in FFT
Sweeps
Since FFT’s do not sweep, you cannot use a Tracking Source
while doing FFT’s
–221 Settings conflict; Trigger
input in use for source
synchronization
If Point Trigger is being used with an external trigger input to
synchronize an external source to the analyzer, that trigger
input is unavailable for triggering.
–221 Settings conflict; Trigger
is not available with span >
0 Hz.
–221 Settings conflict; Tx Band
Spur meas does not support
this frequency band.
The transmit band spur measurement does not support all of
the commercially available frequency bands. You need to
change your selection under Mode Setup, Radio, Band to
one of the supported bands.
–221 Settings conflict; Tx Band
Spur measurement is not
defined for mobiles.
Only base station testing is available.
–221 Settings conflict; Zero Span
not available when Frequency
Scale Type = Log
Logarithmic scales cannot be used for time domain sweeps
(0 Hz span).
–221 Settings
conflict;Administrator
privileges required
You must be logged in with administrator privileges to do
this. Log out and log back in as the administrator, then
restart the SA application.
–221 Settings conflict;Auto Scan
Time/Meas Time do not apply
in Stepped Scan Type.
The Auto Scan Time/Meas Time are not available when Scan
Type =Stepped Scan.
–221 Settings conflict;Auto Scan
Time/Meas Time do not apply
in Stepped Scan Type.
The Auto Scan Time/Meas Time are not available when Scan
Type =Stepped Scan.
–221 Settings conflict;ClearList
& Start not available when
ScanSeq = (Re)measure
Cannot perform Clear List & Start during (Re)measure
because we needs the frequency information of the peaks in
signal list to perform (Re)measure.
–221 Settings conflict;EMI
Detectors and Average
detector can’t be used
together
User is not allowed to turn on any EMI Detector and Average
detector together. They are always mutually exclusive.
–221 Settings conflict;Freq > 1
GHz is not available while
RF Input 2 enabled
Frequency is limited to 1GHz while RF Input 2 is enabled.
Err# Message Verbose/Correction Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals