N9030B PXA Signal Analyzer Service Guide 129
Instrument Messages
Condition Messages
135 1 No Result; Meas invalid with
I/Q inputs
E The current measurement does not
support I/Q input; switch to the RF or
another input or select a different
measurement
137 2 unused
139 3 Uncalibrated Summary status bit
only
This bit is the summary bit for the
STATus:QUEStionable:INTegrity:UNCali
brated sub-register.
141 4 Input Overload W
141 4 Input Overload;ADC over
range
W The signal at the input to the IF section
is too high. You should increase the
attenuation or lower the signal level.
141 4 Input Overload;I/Q ADC over
range
W The I or Q input exceeds the ADC upper
limit.
141 4 Input Overload;I/Q Voltage
over range
W The input voltage on the I or Q channel
exceeds the channel limit. In differential
mode the over voltage may occur
without causing an ADC overload, for
example, if I is at +5.01 V and I-bar is at
+5.0 the ADC will be in range but both I
and I-bar will exceed the voltage limit.
143 5 unused
145 6 unused
147 7 Insufficient Data E
147 7 Insufficient Data; Incr.
Demod Time
E There is insufficient acquisition data to
provide accurate metrics. You should
increase the Demod Time to acquire
enough data.
147 7 Insufficient Data; frequency
list empty
E A measurement was attempted with List
frequency mode or a SCPI query of the
frequency list table was made and the
frequency list table is empty.
147 7 Insufficient Data; ENR table
empty
E A measurement was attempted or a
SCPI query of an ENR table was made
and there were no entries in the relevant
ENR table (Common, Meas or Cal).
147 7 Insufficient Data; Loss
table empty
E A measurement is attempted or a SCPI
query of a before or after loss table is
made and there are no entries in the
relevant loss table
Err# Bit in
status
register
Message Error or
Warning
More Information

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals