134 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
209 3 Timing Error E
209 3 Timing Error:No time
ref pilot burst
E The pilot burst used for time reference is not
active.
211 4 Carrier(s) incorrect
or missing
E In the PSA, this error was reported as one of
the following error numbers: 10165, 10173,
10178, 10419, 10421, 10535, 10560, 10642,
10648, 10650, 10960
213 5 Freq Out of Range E
213 5 Freq Out of Range;
System input (IF)
E One or more system input frequencies are out
of range. If using a frequency list, check that all
entries are valid for current measurement
mode.
213 5 Freq Out of Range;
External LO
E One or more external LO frequencies are out of
range. Check that the LO frequency limits are
set correctly and check the entered
measurement frequencies and measurement
mode.
215 6 Sync Error E W-CDMA: Cannot sync DPCCH pilot.
Cannot synchronize measurement with DPCCH
pilot for Power Control measurement, because
the pilot signal cannot be found. Make sure
DPCCH is present in the W-CDMA uplink
signal, and that the slot format and scramble
code are set correctly.
215 6 Sync Error;No pilot
burst
E There is no Pilot burst detected.
215 6 Sync Error;Sync code
not found
E Synchronization code is not found in the
measured time slot.
215 6 Sync Error;No freq
ref pilot burst
E The pilot burst used for frequency reference is
not active.
215 6 Sync Error;Midamble
sync fail
E Failed to find the uplink slot, which caused the
synchronization with the midamble to fail.
215 6 Sync Error;Preamble
length zero
E Burst type is “Data” or “Preamble” and the
measurement cannot find a Preamble
Err# Bit in
Status
Register
Message Error or
Warning
More Information