136 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
217 7 Demod Error;Not an
active slot
E There is no active slot detected.
217 7 Demod Error;No full
subframe found
E No sub-frame or only part of one sub-frame is
detected.
217 7 Demod Error;Muxed
bits not found
E Multiplexed Data Demod Bits are not
generated even though Data channel is
selected, because all 16 data code channels
are not active
217 7 Demod Error;Acq Time
too short
EFor Bluetooth, the detected packet type
doesn’t match the captured packet type
because the payload start, end or data could
not be found.
219 8 Signal Too Noisy E NADC & PDC: The valid EVM measurement
cannot be performed, because the input signal
is too noisy.
GSM & EDGE: In a GSM measurement,
indicates that a burst could not be found in a
signal that appears noisy.
In the PSA, this error was reported as one of
the following error numbers: 10702, 10824,
10906, 10930, 13024, 10626, 111
221 9 Slot Error E
221 9 Slot error;No active
slot found
E No valid active slot found in captured data, or
no active slot found in captured interval.
Synchronization may succeed and pilot found
when this message is issued, but no results are
included in peak/average calculation.
221 9 Slot Error; No idle
slot found
E No valid idle slot found in captured data, or no
idle slot found in captured interval.
Synchronization may succeed and pilot found
when this message is issued, but no results are
included in peak/average calculation.
223 10 unused E
225 11 unused E
227 12 unused E
229 13 unused E
231 14 unused E
Err# Bit in
Status
Register
Message Error or
Warning
More Information