138 N9030B PXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
309 4 User Cal; Cal invalidated E,W The existing user cal has been invalidated
because of one of the following reasons:
Frequency: Setting the frequency outside
the current valid user cal set (for
example: If the current sweep range is 2
to 3GHz, then setting the start frequency
to 1.9 GHz will invalidate the current user
cal. Other frequency changes that will
invalidate the user cal are:
If the cal was performed at a
fixed frequency and you change
this frequency.
If you are in “Freq List” mode
and you change it to extend
beyond the current user cal
range. In this case you will see
an error message.
DUT Type: If the DUT Type parameter
changes, causing the measurement
frequencies to be pushed outside the
current cal.
Attenuation: If an attenuation setting is
selected but has not been calibrated.
Preamp: If set to condition different from
current cal settings, for example: if
calibrated with the preamp on, turning it
off will invalidate the cal.
Points: Changing the number of
measured frequency points can make the
stored preselector offsets become
inaccurate and hence invalidate the
calibration. This occurs when the
following conditions exist:
A successful calibration has
been performed.
Some measured freq points are
> 3.6 GHz.
The new points > 3.6 GHz are
located more than 50 MHz
away from the current
calibration points.
Err# Status
Register
Bit
Message Error or
Warning
Verbose/Correction Information