PXA RF BLOCK DIAGRAM (Options 503, 508, 513, 526) (Lowband Path - Input signals 3 Hz to 3.599 GHz)
PXA RF (Low Freq)
BLOCK DIAGRAM
A9 INPUT
ATTENUATOR A
A15 FRONT END CONTROLLER
A20 YTO
A13A2 SWITCHED FILTER
A10 INPUT
ATTENUATOR B
A11 LOWBAND
SWITCH
SWITCH 3
(Option LNP)
A12 YTF
SW1
(Option MPB)
SW2
(Option MPB)
A13A1 FRONT END
30 dB
2 dB
2 dB
6 dB
1
1
C
C
2
2
10 dB
-35 dBm
20 dB
6 dB
0 - 24 dB
+20 dB
Cal Signal
From A16 Reference
50 MHz, -25 dBm or
4.8 GHz, -28 dBm
13.6 GHz - 26.5 GHz Path
J7
J9
J8
J13
J1
3.6 GHz - 13.6 GHz Path
3.6 GHz - 26.5 GHz
3 Hz - 3.6 GHz Path
4 GHz
+30 dB
Preamp
Preamp
Electronic Attenuator
Mixer #1
Mixer #2
Mixer #3
1st LO OUT
2nd LO
4800 MHz
+10 dBm
W18
from A16
Reference Bd
260 MHz - 400 MHz
120 MHz
LPF
3.8 - 8.73 GHz
8.3 - 14 GHz
J1
W3
W2
W4
50 MHz
-25 dBm
W8
50 MHz, -35.3 dBm
J900
1,2
AUX
IF OUT
J901
J902
IF In
Ext
IF In
W16
1
IF OUT
322.5 MHz
-22 dBm
EXT IF OUT
W22
J903
J102
J301
J1300
to AIF
322.5 MHz
-23 dBm
W25
W26
Option B40
to Digital IF
Mode = I/Q Analyzer
IF Path 40 MHz
250 MHz, -30 dBm
or
W41
Option B1X
to WB AIF
Mode = I/Q Analyzer
IF Path 140 MHz
300 MHz, -30 dBm
W27
to rear
panel
Option CR3
2nd IF
-30 dBm
Burst
Trigger
J1100
W5
W55
W7
W12
W11
W51 W53
J800
J802
J700
J801
J300 J302
1
4
2
3
W54
W13
W14
W10
Preselector
W15
W52
J3
J3J5
J10
J11
J6
ATTENUATOR CONTROL
LOW BAND SWITCH BYPASS CONTROL
LOW BAND SWITCH CONTROL
PRESELECTOR BYPASS CONTROL
PRESELECTOR DRIVER CONTROL
X2
LOG
DET
J
ALIGNMENT
THERMOMETER
RF
Input
IQ
Modulator
BP F1
BP F2
YTO
Driver
VCO
+10 to
-10 mA
DAC
DAC
LO
DISTRIBUTION
SWITCHED FILTER CONTROL
SWEEP
RAMP
IF OUT from A13A1J7
NOTE: Power levels based on these settings:
Mode: Spectrum Analyzer unless specified differently
RF Calibrator 50 MHz, -25 dBm
Center Frequency 50 MHz
Span 0 Hz
Input Attenuation 10 dB
Preamp OFF
Auto Align OFF after initial Auto Align
AUX IF OUT is only available
with Option CR3, CRP, and ALV
See Low Band troubleshooting
section warning message
Signal levels above 3 GHz include cable loss
of measuring system
FOOTNOTES
1
2
3
IF Path IF Freq.
10 MHz
25 MHz
40 MHz
140 MHz
322.5 MHz
322.5 MHz
250 MHz
300 MHz
BIAS
TEE
MIXER BIAS, AMP BIAS,
PREAMP BIAS, E-ATTEN CONTROL,
SWITCH CONTROL, 1st LO POWER OPTIMIZATION,
YTF TUNE RAMP
W19
W21
W33
W48
W58
W49
W28
J2
J4
W20
IF
Out
Aux LO In
J2
J1
J6
J3 J4
J5
W23
YTO Main and
FM coil current
drive from A14
To A14
Synthesizer
ECAL In
3
IF IN
IF INIF OUT
5172.5 MHz
+10 dBm
5122.5 MHz
-34 dBm
5172.5 MHz
+9 dBm
blockpxa_rf_low
5122.5 zz
-33 dBm
160 MHz BPF
5.1 GHz
SYNTH OUT
5172.5 MHz
+10 dBm
J4
IM
Generator
J12
J2
J2 (Opt EXM)
W29 (Opt EXM)
EXT
MIXER
W17 (Opt EXM)
W30
(Opt EXM)

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals