N9030B PXA Signal Analyzer Service Guide 77
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
AIF 140 Variable Attenuator Algorithm
Ref power at atten 12 dB
(Option B1X)
A25 Wide band A-IF
DIF 40 Variable Attenuator Algorithm
Ref power at atten 21 dB
A3 Digital IF
Narrow Band Step Cal Ad justment Alignment
Algorithm
Alignment signal
originates on A3
D-IF, then goes
through A16
Reference through
the input
attenuators,
through A13 Front
End, through A2
Analog IF, and back
to A3 Digital IF. If
Option B1X present,
A25 WB A-IF
switches calibrator
signal to A16
Reference.
Narrow Band Step Cal Alignment Algorithm A16 Reference
A9/A10 Attenuators
DIF 40 Step Cal Adjustment Alignment
Algorithm
Since this is wide
band alignment
done for 40 MHz
BW signal path, A2
Analog IF is not
included in the
signal path.
DIF 40 Step Cal Alignment Algorithm
AIF 140 Step Cal Adjustment Alignment
Algorithm
(Option B1X)
A25 Wide band A-IF
AIF 140 Step Cal Alignment Algorithm
Table 2-3 Initial Alignments
Alignment Description Most Probable
Hard ware Failure
Related Hard ware

Table of Contents

Other manuals for Keysight Technologies N9030B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N9030B and is the answer not in the manual?

Keysight Technologies N9030B Specifications

General IconGeneral
BrandKeysight Technologies
ModelN9030B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals