N9030B PXA Signal Analyzer Service Guide 91
Instrument Messages
Advisory Messages
LO Fixed freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO fixed freq should be greater than the RF Stop
freq’s for an LSB or DSB (for DSB measurements the setup uses LSB
values) downconverter fixed LO setup. Use the graph icon on the DUT
setup form to clarify the setup required.
LO Start freq should be
greater than IF Fixed freq
The setup frequencies break the rules for an upconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO Start freq should be greater than the IF fixed
freq for an LSB upconverter swept LO setup. Use the graph icon on the
DUT setup form to clarify the setup required.
LO Start freq should be
greater than RF Start freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO start freq should be greater than the RF Start
freq’s for an LSB downconverter swept LO setup. Use the graph icon on
the DUT setup form to clarify the setup required.
LO Stop freq should be
greater than RF Stop freq
The setup frequencies break the rules for a downconverter measurement.
The measurement will still run, but check setup frequencies are correct
before continuing. The LO Stop freq should be greater than the RF Stop
freq’s for a DSB (for DSB measurements the setup uses LSB values)
downconverter swept LO setup. Use the graph icon on the DUT setup form
to clarify the setup required.
Low-pass filter set to OFF Turning on any band pass filter will turn off low-pass filters.
No spurs have been found You has started a measurement in examine meas type in single or
continual sweep mode, or full meas type in single sweep mode, but no
spurs were found.
Preparing Calculation…
Preselector is centered The preselector has been successfully centered
Preselector not used in this
frequency range.
You cannot center or adjust the preselector because it is not used at all at
the current marker frequency or between the current start and stop
frequencies
Probe connected, cal data is
being reapplied; <port>;
<probe>
A probe has been connected, calibration data is being reapplied
Probe connected, no probe
cal; using cable cal data;
<port>; <probe>
A probe has been connected and no probe calibration data is available.
The latest cable calibration data will be used
Probe disconnected,
reverting to cable
calibration data; <port>
A probe has been disconnected, calibration data reverting to the last cable
calibration data
Message Description/Correction Information