Functional Principle, Technical Data
optoNCDT 2300
Model ILD 2310- 10 20 40 50
Measuring range
1
mm
(inches)
10 / 5
(.39 / .20)
20 / 10
(.79 / .39)
40 / 20
(1.57 / .79)
50 / 25
(1.97 / .98)
Start of measuring range
1
mm
(inches)
95 / 100
(3.74 / 3.94)
90 / 100
(3.54 / 3.94)
175 / 195
(6.89 / 7.68)
550 / 575
(21.7 / 22.6)
Midrange
1
mm
(inches)
100 / 102.5
(3.94 / 4.04)
100 / 105
(3.94 / 4.13)
195 / 205
(7.68 / 8.1)
575 / 587.5
(22.6 / 23.1)
End of measuring range
1
mm
(inches)
105 / 105
(4.13 / 4.13)
110 / 110
(4.33 / 4.33)
215 / 215
(8.46 / 8.46)
600 / 600
(23.6 / 23.6)
Measuring rate
7 adjustable stages: 49.14 kHz
2
/ 30 kHz / 20 kHz / 10 kHz / 5 kHz / 2. kHz / 1.5 kHz
Linearity
µm < ± 3 < ± 6 < ± 12 < ± 50
% FSO < ± 0.03 < ± 0.1
Resolution µm 0.5
1
3
2 7.5
Spot diameter
SMR 400 x 500 200 230 400 ... 500
MMR 400 x 500 60 210 400 ... 500
EMR 400 x 500 200 230 400 ... 500
Light source Semiconductor laser < 1 mW, 670 nm (red); Class 2 acc. to IEC 60825-1: 2014 / optional class 3R
Permissable ambient light 10,000…40,000 lx
Power supply 24 VDC (11 ... 30 V); < 3 W (24 V)
Digital interface
RS422 (16 bit) / Ethernet / EtherCAT / PROFINET
4
/ EtherNet/IP
4
Analog output
4
4 … 20 mA / 0 … 5 V / 0 … 10 V / ± 5 V / ± 10 V
Temperature range Operation: 0 ... +50 °C (+32 °F up to +122 °F); storage: -20 ... +70 °C; (non-condensing)
Shock (DIN EN 60068-2-29) 15 g / 6 ms in 3 axis
Vibration (DIN EN 60068-2-6) 2 g / 20 … 500 Hz
Protection class (DIN EN 60529) IP65
Housing size, material M, aluminum L, aluminum
Weight approx. 550 g (incl. pigtail)
FSO = Full Scale Output, SMR = Start of measuring range, MMR = Mid of measuring range, EMR = End of measuring range; The specified data ap-
ply to white, diffuse reflecting surfaces (Micro-Epsilon reference ceramic for ILD sensors)
1) Second value is valid for a measuring rate of 49.14 kHz 3) Measuring rate of 10 kHz, without averaging
2) Measuring rate of 49.14 kHz with reduced measuring range 4) Connection via interface module (see accessories)