272
Troubleshooting
11
11.1. Trouble causes and corrective measures
11.1.2. Weighing operation
Phenomenon Cause Action
Correct
weighing cannot
be executed.
Brand setting has been
incorrectly set.
Correctly set the brand setting
ĺ%UDQGVHWWLQJV
Recipe setting has been
incorrectly set.
&RUUHFWO\VHWWKHUHFLSHVHWWLQJĺ
0HDVXUHPHQWLQUHFLSHPRGH
Weighing
cannot be
executed by
your intended
operation.
)XQFWLRQVUHTXLUHGIRU
each weighing mode have
been incorrectly set.
&RUUHFWO\VHWWKHIXQFWLRQVUHTXLUHGIRUHDFK
weighing mode.
11.1.3. Standard RS-485 interface
Phenomenon Cause Action
Signals are not
transmitted.
The standard RS-485
interface has been
incorrectly wired.
Correctly wire the standard RS-485 interface
ĺ6WDQGDUG56LQWHUIDFHZLULQJ
The address number has
been incorrectly set.
Correctly set the address number
ĺ$GGUHVVVHWWLQJ
The communication
protocol has been
incorrectly set.
Set the communication protocol according to
WKHVSHFL¿FDWLRQVRIWKHKRVWĺ
&RPPXQLFDWLRQVHWWLQJ
Command data have been
incorrectly set.
&RUUHFWO\VHWFRPPDQGGDWDĺ
&RPPDQGPRGHGDWDIRUPDWV
Communication
cannot be
controlled.
7KHFRPPXQLFDWLRQÀRZ
has been controlled by the
host.
0DNHVHWWLQJVVRWKDWWKHFRPPXQLFDWLRQÀRZ
is not controlled on the host side.
The host
responds too
TXLFNO\
The interval from the time
the host receives a
command until it executes
transmission, has been 1
ms or less.
On the host side, set the interval from the time
of command reception to the time of
transmission to 1 ms or more.
The host does
not respond.
The currently set data
delay time for CSD-912B
has been too short.
Set the data delay time of the standard RS-485
interface to a higher value
ĺ'HOD\WLPHVHWWLQJ