4.4.Adjustment after calibration
47
4
Calibration
4.4.6. Calibration lock
You can lock calibration settings to prevent them from being changed. Use the calibration
lock switch inside the CSD-912B to lock the calibration settings.
5HPRYHWKHWZR¿[LQJVFUHZVRIWKHXSSHUFRYHU
2. Slid the upper cover in the direction of the rear panel to remove it.
3. Switch the calibration lock switch to ON.
To release the lock, switch the calibration lock switch to OFF.
POINT
- You cannot change the settings of the following functions while the calibration is
locked.
[CALIBRATION]:
[ZERO ADJUSTMENT], [SPAN ADJUSTMENT], [WEIGHING CAPACITY],
[DIVISION] (minimum scale), [ZERO POINT mV/V], [SPAN POINT mV/V],
[DECIMAL POINT], [UNIT], [BRIDGE POWER SUPPLY], [OL DISPLAY
CONDITION], [ADJ. GRAV. ACCELERATION], [CALIB. PLACE No. ],
[OPERATION PLACE No. ], [CALIB. PLACE G. A. ], [OPERATION PLACE G.
A. ], [FINE ADJUSTMENT], [DIGITAL LINEARISE], [PITCH CONFIRMATION
6(7@>3,7&+&21),50$7,21@ĺ³&DOLEUDWLRQ´³$GMXVWPHQWDIWHU
calibration” “7.1. Settings related to screen display”)
Basic function settings:
[A/D SAMPLING RATE], [OPERATION OF ZERO], [RANGE OF ZERO],
[TARGET OF ZERO TRACKING], [ZERO TRACKING TIME], [ZERO
TRACKING BAND], [OPERATION OF TARE], [RANGE OF TARE], [ZERO AT
POWER ON], [ZERO CLR. AT POWER ON], [TARE AT POWER ON], [TARE
CLR. AT POWER ON], [OPERATION OF STABILITY], [SET “MEASUREMENT
$&7´@ĺ³2SHUDWLRQVGXULQJPHDVXUHPHQW´³6HWWLQJVUHODWHGWR
measurement” “7.2. Settings related to correction”)
Calibration lock switch
Cover tightening
screw
Cover
ON
OFF
3.
2.
1.