USE MANUAL
MINISENTRY-2
107911_D User manual MINISENTRY-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
24 TROUBLESHOOTING
24.1 DAMAGE MOST LIKELY
The following table shows troubleshooting not requiring disassembly of the Minisentry 2
Symptom Action
Minisentry-2 does not start
Connect the mains cable, USB or, depending on the
option, the 12 VDC power supply.
Battery fault, flashing
audible alarm
Replace the internal battery
: Do not replace the battery with a battery or
battery other than that recommended
External battery blinking
Charge the external battery
Replace the external battery
Portal LED flashing purple
Use CSPS to see error details.
2 Frisker LEDs flashing purple
Replace the probe cable
Replace the probe
Difficulty erasing LOG files Format the RAM (See § 9.10.7)