USE MANUAL
MINISENTRY-2
107911_D User manual MINISENTRY-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
The user or vehicle goes through the gate and stops for an undetermined amount of time and measures
until the user leaves.
9.7.1 Before measurement
• The device continuously updates the background noise while no presence is detected.
• The device triggers a blue beacon when the measure is ready to be performed.
• If the background noise is not valid and a presence is detected in the gate, the device displays a
message telling to leave, and the LEDs blinks orange.
• If no presence is detected, the device displays the current background noise, as well as its unit and
its remaining calculation time if applicable.
9.7.2 During measurement
• The device measures until the presence is gone.
• Background noise is subtracted from the measure.
• The device reacts in real time to alarm thresholds and act accordingly.
9.8 FRISKER PROBE OPTION
During either measure mode, a user shall be able to use the external alpha beta probe to perform a
secondary measurement at any time, with both results being displayed at the same time on the screen.
When a FRISKER probe is connected, the LED associated will flash with on the screen the message
« Searching… », « Probe found » then « Initializing »:
Measurement is displayed and the simulated sound becomes on.