USE MANUAL
MINISENTRY-2
98 / 170
107911_D User manual MINISENTRY-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
2
Loches Cedex
13.9 COEFFICIENT CALCULATION
Calculation of the coefficient is no longer systematic.
Verification of efficiency with the default calibration coefficient is sufficient to validate the probe.
A measurement of activity will be proposed to verify that the probe detects the activity of the source
used.
If the measurement is +/- 25% of the source used, the probe must be calibrated with the current date.In
the opposite case, the user has the choice of returning to the high voltage stage and restarting his
calibration or calculating his own calibration coefficient despite the risks.
1. Click on « Calibration » tab: