USER MANUAL
MINISENTRY-2
107911_D User manual MINISENTRY-2_En.docx
MIRION TECHNOLOGIES (Canberra)S.A.S., BP249, ZI de Vauzelles, 37602
Loches Cedex
69 / 170
The SD decision threshold is calculated according to the formula:
=

󰇧
1

+
1

󰇨
With:
is the coefficient of false alarm

is the average value of the BKG

is the time of acquisition of the BKG

is the measurement time
9.11.3 Monitor calibration
9.11.3.1 Check of the calibration date
1. Press the key to access the « Main menu ». and press the or keys to select the
line « Advanced settings ».
2. Press the key to display the « Advanced settings » screen and press the or
keys to select the line « Calibration ».
3. Press the key to display the « Calibration » screen: