USE MANUAL
MINISENTRY-2
107911_D User manual MINISENTRY-2_En.docx
MIRION TECHNOLOGIES (Canberra) S.A.S, BP249, ZI de Vauzelles, 3760
Messages on screen Action
The “High background” alarm option is enabled
and background is upper than the “High
background trip level”.
Adjust the value
With purple portal LED blinking.
In « Enter wait » and « Walk through » modes,
release the portal to perform the background
acquisition
In « Enter wait » mode, the user leaves before
the end of measurement.
« Monitor calibration expired » setting is with
« Warning » option.
The monitor is usable but it is imperative to carry
out a calibration (§ 9.11.3).
With purple LEDs blinking
« Monitor calibration expired » setting is with
« Out of service » option.
The monitor is out of service and it is mandatory
to carry out a calibration (§ 9.11.3).
« Probe calibration expired » setting is with
« Warning » option.
The portal probes is usable but it is imperative to
carry out a calibration with Canberra Smart
Probe Software (CSPS) software (§ Erreur !
Source du renvoi introuvable.)