Chapter E - PREVENTIVE MAINTENANCE
E-2 Planmeca Compact i
INSTRUMENT CONSOLE
Technical Manual
3 INSTRUMENT CONSOLE
Oil collector
Clean the collector.
Instrument arms
Condition of the arms. Replace if needed.
Instrument hoses
Condition of hoses. Replace if needed.
Condition of quick connectors.
Condition of rubber gasket. Replace if needed.
Check for leaks.
Control panel
All the keys are working.
4 INSTRUMENTS
Syringe
Tip is clean.
Turbine
Drive air pressure. Service mode 30.
Bulb voltage.
O-rings. Change automatically twice a year.
•Lubricate.
Micromotor
Bulb voltage.
O-rings. Change automatically twice a year.
•Lubricate.
Polymerization light
Bulb voltage. Recommended voltages are 8.90 Vdc (Lysta), 15.4 Vdc
(Satelec), 10.0 Vdc (Demetron, PowerLysta).