Chapter E - PREVENTIVE MAINTENANCE
E-2 Planmeca Compact i
INSTRUMENT CONSOLE
Technical Manual
3 INSTRUMENT CONSOLE
Oil collector
• Clean the collector.
Instrument arms
• Condition of the arms. Replace if needed.
Instrument hoses
• Condition of hoses. Replace if needed.
• Condition of quick connectors.
• Condition of rubber gasket. Replace if needed.
• Check for leaks.
Control panel
• All the keys are working.
4 INSTRUMENTS
Syringe
• Tip is clean.
Turbine
• Drive air pressure. Service mode 30.
• Bulb voltage.
• O-rings. Change automatically twice a year.
•Lubricate.
Micromotor
• Bulb voltage.
• O-rings. Change automatically twice a year.
•Lubricate.
Polymerization light
• Bulb voltage. Recommended voltages are 8.90 Vdc (Lysta), 15.4 Vdc
(Satelec), 10.0 Vdc (Demetron, PowerLysta).