Chapter D - HELP & ERROR MESSAGES
D-18 Planmeca Compact i
DETAILED DESCRIPTION OF ERROR MESSAGES
Technical Manual
3.7 Foot control related error messages
E 51.n Cable from foot control is loose or connected into wrong
connector on MCB
E 51.1 reserved
E 51.2 Foot control is not responding
The foot control cable is disconnected from the unit, or the cable is open circuited, or the foot
control has failed. Check if the foot control gets power by checking its that its internal yellow
status-LED is blinking. Try to recalibrate the foot control, see page 2.6 “Calibrating foot con-
trol” on page H-15 for details.
E 52 Foot control is of new type, update MAIN software
The recently attached foot control is of a new type, of its software is not compatible with the
MCB software. Please update the MCB software. The new software on the MCB is also still
compatible with the older version(s) of the Prostyle Compact foot controls.
E 53 Foot control data error (cable short circuit)
The data signal is short circuited to ground. Check the condition of the foot control cable and
replace it if necessary. Try with another foot control to determine the location of the problem.
Check also the condition of the internal foot control cable going from the MCB to the foot con-
nector at the rear of the unit base.
E 54.n Foot control calibration error
These errors inform that the foot control calibration failed for some reason. The decimal part
number gives the details:
E 54.1 Foot control EEPROM write failure
E 54.2 Foot control EEPROM read failure
E 54.3 Pedal too far from the sensors, channel A
E 54.4 Pedal too far from the sensors, channel B
E 54.5 Sensor plate is too much tilted to either side, checksum error
E 54.6 Sensor plate is too close to the PCB, channel A
E 54.7 Sensor plate is too close to the PCB, channel B
E 54.8 Foot control re-trigger error.
Calibrate the foot control. If it does not help, check that the capacitance sector
plate on the pedal is not touching the foot control PCB, or is not too close to the
PCB. It is also possible, however very unlikely, that the foot control PCB is
defect.
E 55 reserved

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz

Related product manuals