Chapter F - ADJUSTMENTS
F-32 Planmeca Compact i
WMS OPERATION
Technical Manual
6 WMS OPERATION
6.1 List of abbreviations
o = open
c = closed
Vn = valve number n, refer to the WMS pneumatical diagram
SWn = switch number n, refer to the WMS pneumatical diagram
L1 = upper sensor
L2 = lower sensor
6.2 WMS operation, normal use
Continuously
City Water Bottled Water
SW1 off SW1 on
SW2 on SW2 on
V3 o
exception: c when bottle is filled and
instruments not in use
V3 o
V4 c
exception: o for 150 ms, when unit is
turned on (from software version
3.07)
V4 c
exception: o for 150 ms, when unit is
turned on (from software version
3.07)
V7 o V7 o
Bowl rinse is used, water comes directly from incoming water line.
City Water Bottled Water
Bowl rinse Bowl rinse
V1 o V1 o
V2 c V2 c
V5 c V5 c
V6 o V6 o
Pump not running Pump not running
Display PCI Display PCI