Chapter F - ADJUSTMENTS
Planmeca Compact i F-35
WMS OPERATION
Technical Manual
6.3 WMS operation, WCS use
When door is open, pump is not running and syringe cannot be used.
City Water Bottled Water
Door open Door open
V1 c V1 c
V2 c V2 c
V5 c V5 c
V6 c V6 c
Pump not running Pump not running
Display door Display door
Syringe no water flow Syringe no water flow
When suction is used, the main water valve is open to feed water to suction system.
City Water Bottled Water
Suction in use Suction in use
V1 o V1 o
V4 o for 150 ms (from software version
3.07)
V4 o for 150 ms (from software version
3.07)
When bottle is removed water cannot be used.
City Water Bottled Water
Special situation: bottle removed Special situation: bottle removed
SW2 off SW2 off
V3 c V3 c
V2 c V2 c
V5 c V5 c
Pump not running Pump not running
Display He 62 Display He 62
Continuously
City Water Bottled Water
SW2 on SW2 on
V6 c V6 c
V7 o V7 o