EasyManuals Logo

Planmeca Compact I User Manual

Planmeca Compact I
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
Chapter F - ADJUSTMENTS
F-34 Planmeca Compact i
WMS OPERATION
Technical Manual
Instruments are used, no need to fill bottle.
City Water Bottled Water
Instrument in use (water bottle full) Instrument in use (water bottle full)
V1 c V1 c
V2 c V2 c
V5 c V5 c
V6 c V6 c
Pump not running Pump not running
Display PCI Display PCI
Instruments are used and at same time bottle is filled.
City Water Bottled Water
Instrument in use (water level below upper
sensor)
Instrument in use (water level below upper
sensor)
V1 o V1 c
V2 o V2 c
V5 c V5 c
V6 c V6 c
Pump running Pump not running
Display PCI Display PCI
Instruments are used and at same time bottle is filled.
City Water Bottled Water
Instrument in use (water level below lower
sensor)
Instrument in use (water level below lower
sensor)
V1 o V1 c
V2 o V2 c
V5 c V5 c
V6 c V6 c
Pump running Pump not running
Display He 60 Display He 61

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz

Related product manuals