Chapter F - ADJUSTMENTS
F-34 Planmeca Compact i
WMS OPERATION
Technical Manual
Instruments are used, no need to fill bottle.
City Water Bottled Water
Instrument in use (water bottle full) Instrument in use (water bottle full)
V1 c V1 c
V2 c V2 c
V5 c V5 c
V6 c V6 c
Pump not running Pump not running
Display PCI Display PCI
Instruments are used and at same time bottle is filled.
City Water Bottled Water
Instrument in use (water level below upper
sensor)
Instrument in use (water level below upper
sensor)
V1 o V1 c
V2 o V2 c
V5 c V5 c
V6 c V6 c
Pump running Pump not running
Display PCI Display PCI
Instruments are used and at same time bottle is filled.
City Water Bottled Water
Instrument in use (water level below lower
sensor)
Instrument in use (water level below lower
sensor)
V1 o V1 c
V2 o V2 c
V5 c V5 c
V6 c V6 c
Pump running Pump not running
Display He 60 Display He 61