Chapter B - CONTROL PANEL FUNCTIONS
Planmeca Compact i B-7
PROGRAMMING MODE FUNCTION DETAILS
Technical Manual
4.6 Duration of the bowl rinsing and duration of the glass filling
a) Press the program key.
b) Press the bowl rinse/glass fill key. Press it repeatedly to toggle between bowl rinsing and cup
filling times. The text ‘b.nnn’ indicates bowl rinse time, the text ‘c.nn.n’ indicates cup filling
time.
c) Change the duration with the help of the foot control pedal. Bringing the pedal to the right
increases the time, taking it to the left side will decrease the time.
d) The minimum value of the bowl rinsing is 5 sec. and the maximum 240 sec., and the
adjustment step is 5 seconds. The minimum value of the glass filling is 2 sec. and the
maximum 10 sec., and the adjustment step is 0.5 seconds.
e) Press the program key.
NOTE The duration of the glass filling can alternatively be programmed as follows:
a) Place an empty glass to the glass holder.
b) Press the program key.
c) Press the bowl rinse/glass fill key twice. The text ‘c.nn.n’ will appear on the display.
d) Push and hold the foot control pedal down. The glass filling time will first be set to 2 seconds
and will start to increase after 2 seconds. The pedal can be released and then pushed again,
and the time continues to increase.
e) Press the program key to exit and store the new setting.
b. 30
c. 6.0
+
-

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz