Chapter D - HELP & ERROR MESSAGES
Planmeca Compact i D-7
ERROR MESSAGES
Technical Manual
2.3 Error message short-form table
Error Explanation Details on:
GENERAL
E1.nn
Output (Valve/Aux) short circuit on MCB. page D-10
E2.nn
Output (Valve/Aux) open circuit on MCB. page D-10
E3.n
Internal pressure (water and/or air) problem. page D-11
E4.n
Short circuit in the cable from the MCB to the control panel/suction holder in
branch n.
E5
reserved
E6
reserved
POWER SUPPLY RELATED
E7
IPS (Instrument Power Supply) voltage failure on MCB. page D-11
E8
reserved
E9
Internal operating voltage too low (or mains < 15% under nominal rating). page D-12
E10
Internal operating voltage too high (or mains >15% over nominal rating). page D-12
E11.n
Fuse is blown on MCB. page D-12
E12
reserved
E13
Mains frequency is outside acceptable range (< 47 or >63 Hz). page D-12
E14
IPS heatsink temperature measurement sensor error. page D-12
E15
IPS heatsink running too hot on MCB (heatsink not attached properly to elec-
tronics box).
page D-12
E16.n
Attached scaler electronics is of new type. page D-13
E17
reserved
E18
reserved
CONTROL PANEL RELATED
E19.n
Control panel is of new type, update software on MCB. page D-13
E20.n
Control panel key is stuck during selftest. page D-13
E21
reserved
E22
reserved
INSTRUMENT RELATED
E23
Polymerization (light curing) light bulb is overheated, blown, or not in the
socket.
page D-13
E24
reserved
E25
Satelec mini L.E.D. polymerization light requires IMUX PCB version F or later. page D-13
E26
reserved

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca Compact I and is the answer not in the manual?

Planmeca Compact I Specifications

General IconGeneral
SpittoonYes
Operating LightLED
Imaging TechnologyDigital
X-ray Tube Voltage60-70 kV
X-ray Tube Current7 mA
Focal Spot Size0.8 mm
Exposure Time Range3.2 s
Power Supply50/60 Hz

Related product manuals