EasyManuals Logo

Planmeca Compact I User Manual

Planmeca Compact I
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Chapter F - ADJUSTMENTS
Planmeca Compact i F-29
WATER MANAGEMENT SYSTEM: ADJUSTING THE SEN-
Technical Manual
Pass the door switch by pushing e.g a fork spanner into the electronics control box as
shown on the figure below.
The fork spanner is pressing the door switch and the WMS operates also with door open.