Command reference
R&S
®
ZNA
949User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure:FILename <String>
For a fixture modeling with the ISD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<String> Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
Manual operation: See "Load File" on page 678
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using CALCulate:FMODel:ISD<Ph_pt>:
DUT:MEASure:FILename.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Usage: Event
Manual operation: See "Load File" on page 678
CALCulate:FMODel:ISD<Ph_pt>:DUT:TYPE <DUTType>
Defines whether the DUT is active or passive.
Corresponds to the active_dut batch mode parameter of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<DUTType> PASSive | ACTive
Manual operation: See "DUT Type" on page 683
CALCulate:FMODel:ISD<Ph_pt>:DUT[:STATe] <Boolean>
For a fixture modeling with the ISD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using CALCulate:FMODel:
ISD<Ph_pt>:DUT:MEASure).
Suffix:
<Ph_pt>
.
Physical port number
Parameters:
<Boolean> 1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
SCPI command reference