EasyManua.ls Logo

R&S ZNA67 - Page 950

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
950User Manual 1178.6462.02 ─ 20
Example: See Chapter 8.2.8, "Fixture modeling", on page 1628
Manual operation: See "Active" on page 678
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:DUT <FltDutAndLeadIns>
If CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride is set to TRUE, this com-
mand allows to set the flight time for DUT + Lead-ins manually.
This is equivalent to setting the leadin_dut_time batch mode parameter of the ISD
tool to a numeric value (not auto).
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<FltDutAndLeadIns> Default unit: s
Manual operation: See "Flt Tm for DUT + Lead Ins" on page 683
CALCulate:FMODel:ISD<Ph_pt>:FTIMe:OVERride <Boolean>
If set to true, the flight time for DUT + lead-in is specified using CALCulate:FMODel:
ISD<Ph_pt>:FTIMe:DUT. Otherwise, it is calculated automatically.
True is equivalent to setting the leadin_dut_time batch mode parameter of the ISD
tool to auto.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Boolean>
Manual operation: See "Automatic Flt Tm for DUT + Lead Ins" on page 683
CALCulate:FMODel:ISD<Ph_pt>:OPERation <Operation>
Defines the execution mode of the ISD tool.
Suffix:
<Ph_pt>
.
This suffix is ignored.
Parameters:
<Operation> FAST | ACCurate
ACCurate: normal execution mode
FAST: reduces the execution time to ~50%
In many cases the FAST mode can be enabled with only little
loss of accuracy.
Manual operation: See "Operation (fast/acc)" on page 684