Measurements and Result Displays
R&S
®
ESW
112User Manual 1177.6298.02 ─ 08
Figure 9-8: Overview of the test automation process
For more information on the individual measurement stages, see:
●
Scan
"Scan" on page 107
Chapter 9.3.3, "Performing a Scan", on page 114
●
Peak search
"Peak Search" on page 108
Chapter 9.3.4, "Performing a Peak Search", on page 122
●
Final measurement
"Final Measurement" on page 110
Chapter 9.3.5, "Performing Final Measurements", on page 130
Starting a complete automated test sequence
► Press the [RUN SINGLE] key.
The R&S ESW starts the test sequence. It stops the test sequence at the point you
have defined (either after the scan, the peak search or the final measurement).
Note: If you start the test sequence with the [RUN CONT] key, the R&S ESW starts
a continuous scan. The test sequence is resumed only when you stop the scan
deliberately.
Starting an independent peak search
►
Press the button.
The R&S ESW starts a peak search on the current trace(s) of the scan diagram.
When it is done, it opens the peak list.
If you have not yet performed a scan, the peak search button is unavailable.
Starting an independent final measurement
►
Press the
button.
The R&S ESW starts a final measurement based on the contents of the peak list.
When it is done, it opens the final peak list.
Test Automation