Remote Commands in the Receiver Application
R&S
®
ESW
667User Manual 1177.6298.02 ─ 08
SYSTem:RSWeep <State>
Controls a repeated sweep of the E1 and MKPK HI HP model commands (for details
on the commands refer to Chapter 15.12, "Reference: GPIB Commands of Emulated
HP Models", on page 687). If the repeated sweep is OFF, the marker is set without
sweeping before.
This command is only available if a HP language is selected using SYSTem:LANGuage
on page 664
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYSTem:RSW ON
Manual operation: See "Sweep Repeat" on page 404
15.9.8 System Configuration Check
Commands to check the system configuration described elsewhere.
●
DIAGnostic:SERVice:SINFo? on page 674
DIAGnostic:INFO:CCOunt?.............................................................................................667
DIAGnostic:SERVice:BIOSinfo?...................................................................................... 668
DIAGnostic:SERVice:HWINfo?........................................................................................ 668
DIAGnostic:SERVice:VERSinfo?..................................................................................... 669
SYSTem:ERRor:CLEar:ALL............................................................................................ 669
SYSTem:ERRor:EXTended?........................................................................................... 669
SYSTem:ERRor:LIST?................................................................................................... 670
SYSTem:ERRor[:NEXT]?................................................................................................670
SYSTem:FIRMware:UPDate............................................................................................671
SYSTem:FORMat:IDENt.................................................................................................671
SYSTem:PRESet:COMPatible.........................................................................................671
SYSTem:PRESet:FILTer................................................................................................. 672
SYSTem:SECurity[:STATe]..............................................................................................672
DIAGnostic:INFO:CCOunt? <Relay>
This command queries how many switching cycles the individual relays have per-
formed since they were installed.
Query parameters:
<Relay> ACDC
Mechanical Attenuation Coupling
ATT5
Mechanical Attenuation 05 DB
ATT10
Mechanical Attenuation 10 DB
ATT20
Mechanical Attenuation 20 DB
General Instrument Setup