Remote Commands in the Receiver Application
R&S
®
ESW
715User Manual 1177.6298.02 ─ 08
The STB command and the serial poll respond with an 8-bit value with the following
assignment:
Bit enabled by RQS Description
0 not used (value 0)
1 Units key pressed
2 End of Sweep
3 Device Error
4 Command Complete
5 Illegal Command
6 Service Request
7 not used (value 0)
Bits 0 and 7 are not used and always have the value 0.
Please note that the R&S ESW reports any key pressed on the front panel rather than
only the unit keys if bit 1 was enabled.
Another difference is the behavior of bit 6 when using the STB? query. On the HP ana-
lyzers this bit monitors the state of the SRQ line on the bus. On the R&S ESW this is
not possible. Therefore this bit is set, as soon as one of the bits 1 to 5 is set. It won't be
reset by performing a serial poll.
15.13 Reference: Command Set of Emulated PSA Models
Despite the differences in system architecture and device features, the supported com-
mands have been implemented in a way to ensure a sufficiently high degree of corre-
spondence with the original.
In many cases the selection of commands supported by the R&S ESW is sufficient to
run an existing GPIB program without adaptation.
Supported 89600 commands
*CAL?
*CLS
*ESE
*ESR?
*IDN?
*IST?
*OPC
*OPT?
Reference: Command Set of Emulated PSA Models