Remote Commands in the Receiver Application
R&S
®
ESW
629User Manual 1177.6298.02 ─ 08
15.9.4 Transducers
Before making any changes to a transducer factor or set, you have to select one by
name with [SENSe:]CORRection:TSET:SELect.
Compared to manual configuration of transducers, any changes made to a transducer
factor or set via remote control are saved after the corresponding command has been
sent.
Designing a transducer factor
//Select a transducer factor
CORR:TRAN:SEL 'Transducer1'
//Define a comment for the transducer factor
CORR:TRAN:COMM 'Correction values for device x'
//Define the transducer factor unit
CORR:TRAN:UNIT 'DB'
//Select the scale of the frequency axis
CORR:TRAN:SCAL LOG
//Define the data points of the transducer factor
CORR:TRAN:DATA 1MHZ,-10,10MHZ,-10,100MHZ,-15,1GHZ,-15
//Turn on the transducer
CORR:TRAN ON
//Automatically adjust the reference level
CORR:TRAN:ADJ:RLEV ON
Managing a transducer set
//Select or create a transducer set
CORR:TSET:SEL 'Transducer Set'
//Define a comment for the transducer set
CORR:TSET:COMM 'Transducer set example'
//Define a unit for the transducer set
CORR:TSET:UNIT 'DB'
//Turn the transducer break on
CORR:TSET:BRE ON
//Define the first transducer range
CORR:TSET:RANG1 150KHZ,1MHZ,'Transducer 1','Transducer 3'
CORR:TSET:RANG2 1MHZ,10MHZ,'Transducer 3','Transducer 4'
CORR:TSET:RANG3 10MHZ,30MHZ,'Transducer 3'
//Turn on the transducer set
CORR:TSET ON
Commands to configure transducers described elsewhere.
MMEMory:SELect[:ITEM]:TRANsducer:ALL on page 589
MMEMory:LOAD<n>:TFACtor......................................................................................... 630
MMEMory:STORe<n>:TFACtor....................................................................................... 630