Remote Commands in the Receiver Application
R&S
®
ESW
517User Manual 1177.6298.02 ─ 08
Manual operation: See "Trigger Source" on page 187
See "Free Run" on page 187
See "Ext. Trigger 1/2" on page 187
Table 15-3: Available trigger sources
SCPI parameter Trigger source
EXTernal Trigger signal from the [Trigger Input] connector.
EXT2 | EXT3 Trigger signal from the [Trigger Input/Output] connector.
Note: Connector must be configured for "Input".
IMMediate Free Run trigger.
15.7 Analysis
Result Display Configuration.................................................................................517
Zoomed Displays.................................................................................................. 526
Trace Configuration...............................................................................................528
Markers................................................................................................................. 537
Display and Limit Line Configuration.....................................................................560
15.7.1 Result Display Configuration
General Window Commands................................................................................ 517
Screen Layout.......................................................................................................518
Examples: Configuring the Result Display............................................................524
15.7.1.1 General Window Commands
Note that the suffix <n> always refers to the window in the currently selected measure-
ment channel.
DISPlay:FORMat........................................................................................................... 517
DISPlay[:WINDow<n>]:SIZE........................................................................................... 518
DISPlay:FORMat <Format>
This command determines which tab is displayed.
Parameters:
<Format> SPLit
Displays the MultiView tab with an overview of all active chan-
nels
SINGle
Displays the measurement channel that was previously focused.
*RST: SING
Example:
DISP:FORM SPL