Remote Commands in the Receiver Application
R&S
®
ESW
537User Manual 1177.6298.02 ─ 08
#41024<value1><value2>…<value n> with
#4 Number of digits (= 4 in the example) of the following number of data bytes
1024 Number of following data bytes (= 1024 in the example)
<Value> 4-byte floating point value
Reading out data in binary format is quicker than in ASCII format. Thus, binary format
is recommended for large amounts of data.
15.7.4 Markers
● Individual Marker Configuration............................................................................ 537
● General Marker Configuration...............................................................................541
● Marker Search.......................................................................................................543
● Marker Positioning................................................................................................ 545
● Marker Results......................................................................................................550
● Spectrogram Markers............................................................................................551
15.7.4.1 Individual Marker Configuration
CALCulate<n>:DELTamarker<m>:AOFF...........................................................................537
CALCulate<n>:DELTamarker<m>:MODE......................................................................... 537
CALCulate<n>:DELTamarker<m>:MREFerence................................................................ 538
CALCulate<n>:DELTamarker<m>[:STATe]........................................................................ 538
CALCulate<n>:DELTamarker<m>:TRACe.........................................................................539
CALCulate<n>:DELTamarker<m>:X................................................................................. 539
CALCulate<n>:MARKer<m>:AOFF..................................................................................539
CALCulate<n>:MARKer<m>[:STATe]............................................................................... 540
CALCulate<n>:MARKer<m>:TRACe................................................................................540
CALCulate<n>:MARKer<m>:X........................................................................................ 540
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off all delta markers.
Suffix:
<n>
.
Window
<m> irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:MODE <Mode>
This command defines whether the position of a delta marker is provided as an abso-
lute value or relative to a reference marker.
Analysis