Remote Commands in the Receiver Application
R&S
®
ESW
539User Manual 1177.6298.02 ─ 08
Example:
CALC:DELT2 ON
Turns on delta marker 2.
Manual operation: See "Marker State" on page 231
See "Marker Type" on page 232
See "Select Marker" on page 232
CALCulate<n>:DELTamarker<m>:TRACe <Trace>
This command selects the trace a delta marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Trace> Trace number the marker is assigned to.
Example:
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
CALCulate<n>:DELTamarker<m>:X <Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation: See "Marker Position X-value" on page 231
CALCulate<n>:MARKer<m>:AOFF
This command turns off all markers.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:MARK:AOFF
Switches off all markers.
Manual operation: See "All Markers Off" on page 233
Analysis