89User Manual 1177.6298.02 ─ 08
8.3 Starting an Application
Access
[MODE] > "<application>"
The default application that is running when you start the R&S ESW is the Receiver
application.
The remote commands required to perform these tasks are described in Chapter 15.4,
"Application Selection", on page 435.
The measurement channels are labeled with their default name. If that name already
exists, a sequential number is added.
You can define a different channel name by selecting (double-click) the corresponding
label.
In remote control, the name of the measurement channel can also be changed. For
details and an overview of default names see INSTrument:LIST? on page 437.
Switching between applications
When you enter a new application, a set of parameters is passed on from the current
application to the new one, for example the measurement frequency or the attenuation.
You can also define a fix set of parameters that are synchronized between applica-
tions. For more information see Chapter 13.7, "Synchronizing Measurement Channel
Configuration", on page 347.
Starting an Application