Remote Commands in the Receiver Application
R&S
®
ESW
621User Manual 1177.6298.02 ─ 08
Transducers.......................................................................................................... 629
Display Layout and Elements................................................................................641
Measurement Channel Synchronization............................................................... 648
Network and Remote Control Configuration......................................................... 660
System Configuration Check.................................................................................667
Service Functions..................................................................................................673
15.9.1 Basic Instrument Setup
SYSTem:CLOGging....................................................................................................... 621
SYSTem:REBoot............................................................................................................621
SYSTem:SHUTdown...................................................................................................... 621
SYSTem:CLOGging <State>
This command turns logging of remote commands on and off.
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
Writes all remote commands that have been sent to a file.
The destination is C:\R_S\INSTR\ScpiLogging\
ScpiLog.<no.>.
where <no.> is a sequential number
A new log file is started each time logging was stopped and is
restarted.
OFF | 0
*RST: 0
Manual operation: See "I/O Logging" on page 402
SYSTem:REBoot
This command reboots the instrument, including the operating system.
SYSTem:SHUTdown [<Unit>]
Performs a shutdown or restart of the FW or OS.
If the optional parameter <Unit> is omitted, Windows is shutdown afer a time-out
period of 10 seconds.
Setting parameters:
<Unit> HALT | REBoot | ABORt | CLOSe | RESTart
HALT
Windows is shutdown afer a time-out period of 20 seconds
REBoot
Windows is restarted afer a time-out period of 20 seconds
General Instrument Setup