9 Measurements and Result Displays
The measurements and result displays available in the Receiver application allow you
to analyze the frequency spectrum for possible interferers and the characteristics of
those interferers.
● Performing Measurements......................................................................................95
● Bargraph Configuration...........................................................................................97
● Test Automation.................................................................................................... 102
● CISPR APD Measurements..................................................................................137
9.1 Performing Measurements
The R&S ESW provides several approaches to do measurements and acquire the sig-
nal data.
Access: [SWEEP]
Functions in the "Sweep" menu described elsewhere:
●
"Measurement Time" on page 100
●
"Selecting the scan type" on page 113
●
"Select Frame" on page 216
●
Chapter 9.3.3.1, "Designing a Scan Table", on page 114
Performing continuous measurements......................................................................... 95
Performing single measurements................................................................................. 96
Performing sequenced measurements......................................................................... 96
Performing continuous measurements
When you measure continuously, the measurement is repeated until you deliberately
stop it.
The duration of each individual measurement cycle (for example a single bargraph
measurement) depends on the measurement time you have set for the measurement
in question. After each measurement cycle, the data of the previous cycle is overwrit-
ten and replaced by the new data. Exception: When you are using a max or min hold
function, the application keeps the highest or lowest values and replaces old data only
if the current data contains a new maximum or minimum.
The R&S ESW allows you to select the measurement mode for the bargraph and the
scan separately. When you select a continuous bargraph measurement, IF analysis
(including the spectrogram) is also performed continuously. When you select a continu-
ous scan, the scan spectrogram is also updated continuously. Running a continuous
final measurement is not possible.
▶ To start a continuous scan, press either the [RUN CONT] key on the front panel or
the "Continuous Scan" softkey in the "Sweep" menu.
▶ To start a continuous bargraph measurement, press the "Continuous Bargraph" soft-
key in the "Sweep" menu.