Remote Commands in the Receiver Application
R&S
®
ESW
617User Manual 1177.6298.02 ─ 08
15.8.7 Measurement Result Export
Commands to export measurement results described elsewhere.
FORMat[:DATA] on page 580
MMEMory:STORe<n>:SGRam........................................................................................617
MMEMory:STORe<n>:SPECtrogram............................................................................... 617
MMEMory:STORe<n>:TRACe.........................................................................................617
MMEMory:STORe<n>:SGRam <FileName>
MMEMory:STORe<n>:SPECtrogram <FileName>
This command exports spectrogram data to an ASCII file.
The file contains the data for every frame in the history buffer. The data corresponding
to a particular frame begins with information about the frame number and the time that
frame was recorded.
Note that, depending on the size of the history buffer, the process of exporting the data
can take a while.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see Chapter 4.7, "Protecting Data Using the Secure User Mode",
on page 40.
Suffix:
<n>
.
Window
Parameters:
<FileName> String containing the path and name of the target file.
Example:
MMEM:STOR:SGR 'Spectrogram'
Copies the spectrogram data to a file.
MMEMory:STORe<n>:TRACe <Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
For details on the file format see Chapter 11.3.8.1, "Reference: ASCII File Export For-
mat", on page 224.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256 MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.