Remote Commands in the Receiver Application
R&S
®
ESW
484User Manual 1177.6298.02 ─ 08
CALCulate:MARKer:MAXimum
CALC:MARK:Y?
//Change sweep time in IQ
SENS:SWE:TIME 300us
//Switch to single Sequencer mode
INITiate:SEQuencer:MODE SINGle
//Sweep all channels once, taking the sweep count in each channel into account
INITiate:SEQuencer:IMMediate;*OPC?
//Set marker to maximum in IQ1 and query result
CALCulate:MARKer:MAXimum
CALC:MARK:Y?
//Switch to second IQ channel and retrieve results
INST:SEL 'IQ 2';*WAI
CALCulate:MARKer:MIN
CALC:MARK:Y?
//Switch to first Spectrum channel
INST:SEL 'Spectrum';*WAI
//Query one of the SEM results
CALCulate:MARKer:FUNCtion:POWer:RESult? CPOWer
//Switch to second Spectrum channel
INST:SEL 'Spectrum 2';*WAI
//Query channel power result
CALCulate:MARKer:FUNCtion:POWer:RESult? ACPower
15.6 Configuration
● Input Configuration................................................................................................484
● Output Configuration.............................................................................................495
● Frequency Configuration.......................................................................................503
● Amplitude Configuration........................................................................................506
● Diagram Scale.......................................................................................................510
● Bandwidth and Filter Configuration....................................................................... 511
● Trigger Configuration............................................................................................ 513
15.6.1 Input Configuration
● RF Input Configuration..........................................................................................484
● Preselector Configuration......................................................................................486
● External Generator Control (Optional).................................................................. 487
● User Port Configuration (AUX Port)...................................................................... 493
15.6.1.1 RF Input Configuration
INPut<ip>:COUPling.......................................................................................................485
INPut<ip>:IMPedance.....................................................................................................485
INPut<ip>:TYPE............................................................................................................ 485
Configuration