Remote Commands in the Receiver Application
R&S
®
ESW
454User Manual 1177.6298.02 ─ 08
Example: //Define measurement time
SCAN:TDOM 100 s
Manual operation: See "Measurement Time" on page 100
[SENSe:]FREQuency:TDOPtim <State>
This command selects the mode for time domain scans.
Parameters:
<State> AUTO
Effects depend on selected detectors.
Measurements with CISPR detectors: Automatic mode applies a
small analysis bandwidth (FFT size) in favor of a high dynamic
range.
Measurements without CISPR detectors: Automatic mode
applies a large analysis bandwidth (FFT size) in favor of an
increased measurement speed.
DYNamic
Always applies a small analysis bandwidth (FFT size) in favor of
a high dynamic range, regardless of the selected detectors.
*RST:
AUTO
Example: //Apply small FFT size
FREQ:TDOP DYN
Manual operation: See "Selecting the mode for time domain scans" on page 114
15.5.6 Scan Table Configuration
[SENSe:]DETector<t>[:FUNCtion]....................................................................................455
[SENSe:]FREQuency:STARt........................................................................................... 455
[SENSe:]FREQuency:STOP............................................................................................456
[SENSe:]SCAN<sr>:BANDwidth:RESolution.....................................................................456
[SENSe:]SCAN<sr>:BARS..............................................................................................456
[SENSe:]SCAN<sr>:INPut:ATTenuation:AUTO..................................................................457
[SENSe:]SCAN<sr>:INPut:ATTenuation[:VALue]............................................................... 457
[SENSe:]SCAN<sr>:INPut:GAIN:AUTO............................................................................457
[SENSe:]SCAN<sr>:INPut:GAIN:LNA:AUTO.....................................................................458
[SENSe:]SCAN<sr>:INPut:GAIN:LNA:STATe.................................................................... 458
[SENSe:]SCAN<sr>:INPut:GAIN:STATe........................................................................... 458
[SENSe:]SCAN<sr>:INPut:TYPE..................................................................................... 459
[SENSe:]SCAN<sr>:NAME............................................................................................. 459
[SENSe:]SCAN<sr>:RANGes[:COUNt].............................................................................459
[SENSe:]SCAN<sr>:STARt............................................................................................. 460
[SENSe:]SCAN<sr>:STEP.............................................................................................. 460
[SENSe:]SCAN<sr>:STOP..............................................................................................461
[SENSe:]SCAN<sr>:TIME...............................................................................................461
[SENSe:]SWEep:COUNt.................................................................................................461
[SENSe:]SWEep:SPACing.............................................................................................. 462
Measurements and Result Displays