Remote Commands in the Receiver Application
R&S
®
ESW
459User Manual 1177.6298.02 ─ 08
Example: //Turn on preamplifier for 4th scan range.
SCAN4:INP:GAIN:STAT ON
Manual operation: See "Preamplifier" on page 121
[SENSe:]SCAN<sr>:INPut:TYPE <Input>
This command selects the RF input used for a scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Input> INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST: INPUT1
Example: //Select RF input path for the 4th scan range
SCAN4:INP:TYPE INPUT2
Manual operation: See "Input Selection" on page 122
[SENSe:]SCAN<sr>:NAME <Name>
This command changes the name of a scan range.
Suffix:
<sr>
.
Selects the scan range.
Parameters:
<Name> String containing the name of the scan range.
Example: //Change name of scan range 4
SCAN4:NAME 'Fourth Range'
Manual operation: See "Configuring scan ranges" on page 118
[SENSe:]SCAN<sr>:RANGes[:COUNt] <Range>
This command defines the number of scan ranges.
Suffix:
<sr>
.
irrelevant