EasyManua.ls Logo

Rohde & Schwarz ESW Series - Remote Errors

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Network and Remote Operation
R&S
®
ESW
407User Manual 1177.6298.02 ─ 08
LXI Manufacturer Description
Instrument description of the R&S ESW
Remote command:
SYSTem:LXI:MDEScription on page 665
LAN Reset
Resets the LAN configuration to its default settings (LCI function).
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter Value
TCP/IP Mode DHCP + Auto IP Address
Dynamic DNS Enabled
ICMP Ping Enabled
Password for LAN configuration LxiWebIfc
The LAN settings are configured in the "Network" tab of the "Network + Remote" dialog
box or using the instrument's LXI Browser interface.
Remote command:
SYSTem:LXI:LANReset on page 665
14.4.5 Remote Errors
Access: [SETUP] > "Network + Remote" > "Remote Errors " tab
The error messages generated by the R&S ESW during remote operation are dis-
played here.
The messages are displayed in the order of their occurrence; the most recent mes-
sages are placed at the top of the list.
Network and Remote Control Settings

Table of Contents

Related product manuals