Remote Commands in the Receiver Application
R&S
®
ESW
510User Manual 1177.6298.02 ─ 08
Parameters:
<State> ON | OFF | 1 | 0
*RST: OFF
Example: //Turn on preamplifier
INP:GAIN:STAT ON
Manual operation: See "Preamplifier" on page 121
15.6.5 Diagram Scale
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]................................................................... 510
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:BOTTom......................................................510
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe] <Range>
This command defines the display range of the y-axis (for all traces).
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Range> Range: 1 dB to 200 dB
*RST: 100 dB
Default unit: HZ
Example:
DISP:TRAC:Y 110dB
Manual operation: See "Grid Range / Minimum Level" on page 175
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:BOTTom <Level>
This command defines the minimum level displayed on the vertical diagram axis.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Level> Minimum displayed level. The unit and value range depend on
the unit you have selected.
*RST: 0 dBµV
Example: //Define a minimum grid level
DISP:TRAC:Y:BOTT -20
Manual operation: See "Grid Range / Minimum Level" on page 175