EasyManua.ls Logo

Rohde & Schwarz ESW Series - Frequency Configuration

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
503User Manual 1177.6298.02 ─ 08
Suffix:
<tp>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Usage: Event
Manual operation: See "Send Trigger" on page 169
OUTPut:TRIGger<tp>:PULSe:LENGth <Length>
This command defines the length of the pulse generated at the trigger output.
Suffix:
<tp>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Length> Pulse length in seconds.
Default unit: S
Example:
OUTP:TRIG2:PULS:LENG 0.02
Manual operation: See "Pulse Length" on page 169
15.6.3 Frequency Configuration
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]....................................................503
CALCulate<n>:MARKer<m>:FUNCtion:STRack:BANDwidth.............................................. 504
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold...............................................504
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe.................................................... 504
DISPlay[:WINDow<n>]:TRACe<t>:X:SPACing.................................................................. 505
[SENSe:]FREQuency:CENTer......................................................................................... 505
[SENSe:]FREQuency:CENTer:STEP................................................................................506
[SENSe:]FREQuency:SCOupled..................................................................................... 506
[SENSe:]FREQuency:SPAN............................................................................................506
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe] <State>
This command turns signal tracking on and off.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Configuration

Table of Contents