Remote Commands in the Receiver Application
R&S
®
ESW
580User Manual 1177.6298.02 ─ 08
15.8.1 File Management
FORMat[:DATA]............................................................................................................. 580
FORMat:DEXPort:DSEParator........................................................................................ 581
MMEMory:CATalog.........................................................................................................581
MMEMory:CATalog:LONG...............................................................................................582
MMEMory:CDIRectory....................................................................................................582
MMEMory:COMMent......................................................................................................582
MMEMory:COPY........................................................................................................... 582
MMEMory:DATA............................................................................................................ 583
MMEMory:DELete:IMMediate..........................................................................................583
MMEMory:MDIRectory................................................................................................... 583
MMEMory:MOVE...........................................................................................................584
MMEMory:MSIS.............................................................................................................584
MMEMory:NAME........................................................................................................... 584
MMEMory:NETWork:DISConnect.................................................................................... 584
MMEMory:NETWork:MAP...............................................................................................585
MMEMory:NETWork:UNUSeddrives................................................................................ 585
MMEMory:NETWork:USEDdrives....................................................................................585
MMEMory:RDIRectory....................................................................................................585
FORMat[:DATA] <Format>[, <BitLength>]
This command selects the data format that is used for transmission of trace data from
the R&S ESW to the controlling computer.
Note that the command has no effect for data that you send to the R&S ESW. The
R&S ESW automatically recognizes the data it receives, regardless of the format.
For details on data formats see Chapter 15.7.3.6, "Formats for Returned Values: ASCII
Format and Binary Format", on page 536.
Parameters:
<Format> ASCii | REAL | UINT | MATLab
ASCii
ASCii format, separated by commas.
This format is almost always suitable, regardless of the actual
data format. However, the data is not as compact as other for-
mats may be.
REAL
Floating-point numbers (according to IEEE 754) in the "definite
length block format".
The format setting REAL is used for the binary transmission of
trace data.
<BitLength>
16 | 32 | 64
Length in bits for floating-point results
16
16-bit floating-point numbers.
Compared to REAL,32 format, half as many numbers are
returned.